Dongguan Right Testing Machine Co.,Ltd is manufacturer of Moving Die Rheometer,In order to ensure the normal operation of the Rotorless Rheometer, to extend the service life,Dongguan Right Testing Machine Co.,Ltd sum up some of the maintenance and precautions to everyone, users must do following maintenance,please.
I.Moving Die Rheometer mechanical parts
1.The inside and outside of the instrument must be kept clean. Do not use corrosive solution or gasoline to clean the instrument surface.
2.Every one or two weeks use a soft cloth with oil cleaning two guide post; every month add oil to the bearing, add lubricating oil wear section; if no use for a long time, in the upper and lower mold surface coated with lubricating oil;
3.Each time to finish the test, clean rubber residue in the mold cavity, to keep the mold cavity cleaning
4.No Rotor Rheometer pneumatic parts should be use kerosene clean after working every 3 months.
5.Pressure gauges identified every 6 months
6.The atomized lubricator should be kept oiled and added to 2/3 height.
II.No Rotor Rheometer electrical parts
1.If Moving Die Rheometer failure, first check whether the fuse broken, if broken, should first analysis of what causes, to find out why,then change new fuse.
2.When the fuses are intact, and then check whether the connector loosen.
3.Check whether the switching power supply meets the nominal value, that is, the output should be +5 V, +12 V, if any abnormal, immediately turn off power.
4.The repair principle of circuit board adopt replace the whole board method, the user can purchase a rotorless rheometer with separately board, the warranty period is strictly prohibited modification , or at your peril!
5.The consumable parts of the printer should be changed frequently.
6.Moving Die Rheometer must be energized once a week if long time no use ,charge time not more than 15 minutes, in order to keep the electronic components in a normal state.
7.Circuit board, testing control module users generally do not split, there should be maintenance once have problems.
8.Do not touch the integrated circuit chip on the printed circuit board to avoid electrostatic breakdown!
For more Moving Die Rheometer related questions,please call our hotline:+86 132 6746 6269 or visit http://www.dgright.com
I.Moving Die Rheometer mechanical parts
1.The inside and outside of the instrument must be kept clean. Do not use corrosive solution or gasoline to clean the instrument surface.
2.Every one or two weeks use a soft cloth with oil cleaning two guide post; every month add oil to the bearing, add lubricating oil wear section; if no use for a long time, in the upper and lower mold surface coated with lubricating oil;
3.Each time to finish the test, clean rubber residue in the mold cavity, to keep the mold cavity cleaning
4.No Rotor Rheometer pneumatic parts should be use kerosene clean after working every 3 months.
5.Pressure gauges identified every 6 months
6.The atomized lubricator should be kept oiled and added to 2/3 height.
II.No Rotor Rheometer electrical parts
1.If Moving Die Rheometer failure, first check whether the fuse broken, if broken, should first analysis of what causes, to find out why,then change new fuse.
2.When the fuses are intact, and then check whether the connector loosen.
3.Check whether the switching power supply meets the nominal value, that is, the output should be +5 V, +12 V, if any abnormal, immediately turn off power.
4.The repair principle of circuit board adopt replace the whole board method, the user can purchase a rotorless rheometer with separately board, the warranty period is strictly prohibited modification , or at your peril!
5.The consumable parts of the printer should be changed frequently.
6.Moving Die Rheometer must be energized once a week if long time no use ,charge time not more than 15 minutes, in order to keep the electronic components in a normal state.
7.Circuit board, testing control module users generally do not split, there should be maintenance once have problems.
8.Do not touch the integrated circuit chip on the printed circuit board to avoid electrostatic breakdown!